Home
if (!lastItem) {
>
}
Uranium: Mining, Conversion & Enrichment
if (!lastItem) {
>
}
Sources for dose calibrator quality control
LEA supplies sealed check sources for the verification of dose calibrators used in nuclear medicine.
1
Challenges
Perform the regulatory constancy, repeatability and accuracy checks required for dose calibrators
Have the two check sources required to carry out these controls
Meet the minimum activity requirements for the selected radionuclides
Use sources supplied with a calibration certificate
Handle the sources under conditions suited to verification operations
2
Our Solution
Rely on sealed sources packaged in plastic vials for easier handling
Benefit from a homogeneous distribution of the radionuclide within a thermosetting resin
Use radionuclides with emission lines close to those of
¹⁸F, ⁹⁹ᵐTc, ¹²³I, ¹³¹I, ¹¹¹In
and
²⁰¹Tl
, which are commonly used in nuclear medicine
Select the appropriate reference from LEA’s range of
¹³³Ba, ⁵⁷Co and ¹³⁷Cs
sources
Easy-to-handle sealed sources
Homogeneous radionuclide distribution within the resin
3
Key Features
The standard type F geometry has a useful
volume of 1
0 cm³,
a
diameter of 24 mm
and
a
height of 58 mm
The
¹³³Ba
sources are available with
an activity of 10 MBq
under reference
BA133 EGA F 50
, or
20 MBq under reference BA133 EGA F 55
The
⁵⁷Co
sources are available with
an activity of 100 MBq
under reference
CO57 EGA F 80
,
200 MBq under reference CO57 EGA F 90
, or
370 MBq under reference CO57 EGA F 100
The
¹³⁷Cs
sources are available with
an activity of 10 MBq
under reference
CS137 EGA F 50, or 20 MBq under reference CS137 EGA F 55
At the time of testing, the minimum required activities are
40 MBq for ⁵⁷Co and 5 MBq for ¹³³Ba and ¹³⁷Cs
Upon request
, calibration can be performed under Cofrac accreditation
, with a measurement uncertainty of
3%
or less for a coverage factor of
k = 2
4
Useful Links
Orano LEA Website
LEA Catalog
Regulatory document
Product Sheet : Dose Calibrator Sources